Practical Laboratory Medicine (Aug 2024)
Analytical performance evaluation of a new integrated clinical chemistry and immunoassay analyzer
Abstract
Background: Clinical laboratories perform a wide range of tests that are used by healthcare professionals to guide medical decision making. Use of automated analyzers in the clinical laboratory can improve patient care by not only reducing the turn-around-time (TAT) of results but also improving accuracy of the reported results by reducing human error. The aim of this study was to evaluate the performance characteristics of a new automated laboratory instrument, the Atellica® CI Analyzer, Model 1900, over a 3-month period in a European laboratory setting. Methods: Analytical performance of 17 analytes (13 chemistry and four immunochemistry) was assessed by evaluating repeatability and within-laboratory precision using anonymized remnant serum samples. Method comparison studies were performed on the Atellica CI Analyzer and the Roche cobas® 6000. Results: Excellent precision was observed with coefficients of variation (CVs) less than 2 % for repeatability and less than 3 % within-laboratory imprecision for most analytes. Comparison of select assays with the cobas 6000 system resulted in correlation coefficients ranging from 0.980 to 1.000. Conclusion: This is the first reported evaluation of the Atellica CI Analyzer in a clinical laboratory setting. The strong analytical performance of the Atellica CI Analyzer demonstrates that this instrument is suitable for routine clinical use.