Journal of Orthopaedic Surgery (Dec 2008)

Prosthesis Design and Stress Profile after Hip Resurfacing: A Finite Element Analysis

  • A Heijink,
  • ME Zobitz,
  • R Nuyts,
  • BF Morrey,
  • KN An

DOI
https://doi.org/10.1177/230949900801600312
Journal volume & issue
Vol. 16

Abstract

Read online

Purpose. To evaluate the effect of prosthesis design on stress profile in the proximal femur after hip resurfacing. Methods. The von Mises stress profile of the native femur was simulated and compared with that of resurfaced femurs using various prosthetic materials (titanium, cobalt-chrome, ceramic), stem lengths (normal, half, short, and no stem), and femoral head coverage (shell size) [260°, 220°, 180°, and 140°]. Results. Hip resurfacing altered the stress profile of the cancellous (but not cortical) bone of the femoral neck. Maximal cortical stresses were observed at the posterior half of the medial femoral neck. The stress profile of the native femur was most similar to that of the resurfaced femur made of titanium, with a short or no stem and 260 degrees of femoral head coverage (shell size). Conclusion. Optimising prosthesis design by minimising biomechanical alterations seems a valid approach to achieving favourable long-term outcomes. Cadaveric and in vivo studies are needed to confirm the clinical relevance and feasibility.