Trends in Pharmaceutical Sciences (Mar 2019)

Concurrent analysis of Simvastatin and citicoline using a Reversed-phase High Performance Liquid Chromatography-Ultra Violet Method

  • Negin Mozafari,
  • Soha Azadi,
  • Soliman Mohammadi Samani,
  • Fatemeh Farjadian,
  • Amir Azadi

DOI
https://doi.org/10.30476/tips.2019.81749.1006
Journal volume & issue
Vol. 5, no. 1
pp. 41 – 46

Abstract

Read online

To develop and evaluate a sensitive, accurate, rapid and reproducible high performance liquid chromatography analytical method for concurrent assay of simvastatin, a hyperlipidemia controlling agent, and citicoline, a psychostimulant agent, a C18 column (Eurosphar 100-5, 150 mm × 4.6 mm) used as a reversed stationary phase and mobile phase was water (previously adjusted with phosphoric acid to a pH of 5.5), methanol and acetonitrile (20:20:60) with the flow rate 1.0 ml/min. The ultraviolet detector was set at 247 nm. A linear correlation between each concentration and its own AUC within concentration ranges of 15 to 100 g/ml for citicoline and 7.5 to 50 g/ml for simvastatin with a correlation coefficient 0.9969 for citicoline and 0.994 for simvastatin were produced. The within and between-day precision and accuracy were both in acceptable ranges. The outcomes of these tests show an accurate, rapid and robust HPLC-UV method for successful analysis of both simvastatin and citicoline simultaneously.