Vision Pan-America (May 2015)

Avaliação da Reprodutibilidade das Medições da Espessura da Camada de Fibras Nervosas Peripapilar usando o OCT de Domínio Espectral

  • Ines Coutinho,
  • Mafalda Mota,
  • Catarina Pedrosa,
  • Mário Ramalho,
  • Cristina Santos,
  • Fernando Vaz

DOI
https://doi.org/10.15324/vpa.v14i2.254
Journal volume & issue
Vol. 14, no. 2
pp. 46 – 50

Abstract

Read online

Purpose: To assess the reproducibility of peripapillary retinal nerve fiber layer (RNFL) thickness, measured with Spectralis® OCT, in healthy and glaucomatous subjects, through of inter-examiner variability, intra-test and inter-test with and without software of followup. Methods: Cross-sectional and prospective study with 37 controls (63 eyes) and 33 patients with glaucomatous optic neuropathy (55 eyes). Intraclass correlation coefficient (ICC) and coefficient of variation (CV) were calculated. Results: Low inter-examiner, intra-test and inter-test variability, with ICC>0,75 e CV<10%, in both control group and patients with glaucoma. Conclusions: The Spectralis® OCT shows excellent reproducibility of RNFL measurements, validating it as an useful tool for monitoring the progression of glaucoma.

Keywords