Radiation Oncology (Jul 2020)

Study on the ability of 3D gamma analysis and bio-mathematical model in detecting dose changes caused by dose-calculation-grid-size (DCGS)

  • Han Bai,
  • Sijin Zhu,
  • Xingrao Wu,
  • Xuhong Liu,
  • Feihu Chen,
  • Jiawen Yan

DOI
https://doi.org/10.1186/s13014-020-01603-6
Journal volume & issue
Vol. 15, no. 1
pp. 1 – 11

Abstract

Read online

Abstract Objective To explore the efficacy and sensitivity of 3D gamma analysis and bio-mathematical model for cervical cancer in detecting dose changes caused by dose-calculation-grid-size (DCGS). Methods 17 patients’ plans for cervical cancer were enrolled (Pinnacle TPS, VMAT), and the DCGS was changed from 2.0 mm to 5.0 mm to calculate the planned dose respectively. The dose distribution calculated by DCGS = 2.0 mm as the “reference” data set (RDS), the dose distribution calculated by the rest DCGS as the“measurement”data set (MDS), the 3D gamma passing rates and the (N) TCPs of the all structures under different DCGS were obtained, and then analyze the ability of 3D gamma analysis and (N) TCP model in detecting dose changes and what factors affect this ability. Results The effect of DCGS on planned dose was obvious. When the gamma standard was 1.0 mm, 1.0 and 10.0%, the difference of the results of the DCGS on dose-effect could be detected by 3D gamma analysis (all p value 0.05 accounted for the majority. With DCGS = 2.0 mm being RDS, ∆gamma-passing-rate presented the same trend with ∆(N) TCPs of all structures except for the femurs only when the 1.0 mm, 1.0 and 10.0% standards were adopted for the 3D gamma analysis. Conclusions The 3D gamma analysis and bio-mathematical model can be used to analyze the effect of DCGS on the planned dose. For comparison, the former’s detection ability has a lot to do with the designed standard, and the latter’s capability is related to the parameters and calculated accuracy instrinsically.

Keywords