JACC: Case Reports (Sep 2023)

First Case Report of Electromagnetic Interference Between Percutaneous Ventricular Assist Device and Implantable Cardioverter-Defibrillator

  • Kosuke Takahari, MD,
  • Takafumi Oka, MD, PhD,
  • Takayuki Sekihara, MD,
  • Kentaro Ozu, MD,
  • Yasuhiro Akazawa, MD, PhD,
  • Daisuke Nakamura, MD,
  • Isamu Mizote, MD, PhD,
  • Tomohito Ohtani, MD, PhD,
  • Shungo Hikoso, MD, PhD,
  • Yasushi Sakata, MD, PhD

Journal volume & issue
Vol. 21
p. 101981

Abstract

Read online

Electromagnetic interference (EMI) between implantable left ventricular assist devices and cardiac implantable electronic devices has been observed. We demonstrated the first case of EMI between a percutaneous ventricular assist device and an implantable cardioverter-defibrillator, validated by an extra vivo simulation test. EMI might depend on the distance between devices. (Level of Difficulty: Advanced.)

Keywords