Diagnostics (Nov 2021)

How Reproducible Is the Data from Sysmex DI-60 in Leukopenic Samples?

  • Sumi Yoon,
  • Mina Hur,
  • Gun Hyuk Lee,
  • Minjeong Nam,
  • Hanah Kim

DOI
https://doi.org/10.3390/diagnostics11122173
Journal volume & issue
Vol. 11, no. 12
p. 2173

Abstract

Read online

Digital morphology (DM) analyzers are widely applied in clinical practice. It is necessary to evaluate performances of DM analyzers by focusing on leukopenic samples. We evaluated the analytical performance, including precision, of a Sysmex DI-60 system (Sysmex, Kobe, Japan) on white blood cell (WBC) differentials in leukopenic samples. In a total of 40 peripheral blood smears divided into four groups according to WBC count (normal, mild, moderate, and severe leukopenia; each group n = 10), we evaluated precision of WBC preclassificaiton by DI-60. %coefficients of variation (%CVs) of precision varied for each sample and for each cell class; the fewer cells per slide, the higher %CV. The overall specificity and efficiency were high for all cell classes except plasma cells (95.9–99.9% and 90.0–99.4%, respectively). The largest absolute value of mean difference between DI-60 and manual count in each group was: 10.77, normal; 10.22, mild leukopenia; 19.09, moderate leukopenia; 47.74, severe leukopenia. This is the first study that evaluated the analytical performance of DI-60 on WBC differentials in leukopenic samples as the main subject. DI-60 showed significantly different performance depending on WBC count. DM analyzers should be evaluated separately in leukopenic samples, even if the overall performance was acceptable.

Keywords