International Journal of Ophthalmology (Jan 2016)

Comparison of intraocular pressure measurements with different contact tonometers in young healthy persons

  • Saulius Galgauskas,
  • Rasa Strupaite,
  • Ernesta Strelkauskaite,
  • Rimvydas Asoklis

DOI
https://doi.org/10.18240/ijo.2016.01.13
Journal volume & issue
Vol. 9, no. 1
pp. 76 – 80

Abstract

Read online

AIM: To analyze the correlation of Goldmann applanation tonometer (GAT), I-Care tonometer and Tono-Pen tonometer results in young healthy persons, and to investigate the influence of central corneal thickness (CCT) on intraocular pressure (IOP) measurements recorded with these tonometers. METHODS: We conducted a pilot clinical study in 78 eyes of 78 subjects aged 22-28 years old (44 women and 34 men; mean age 23.8±1.19y). IOP was measured using GAT, I-Care and Tono-Pen tonometers, followed by measurements of CCT. Statistical analysis was performed using SPSS 20.0. RESULTS: The mean IOPs and standard deviation (±SD) for GAT, I-Care and Tono-Pen were 15.62±2.281 mm Hg, 16.29±2.726 mm Hg and 16.32±2.393 mm Hg, respectively. The mean CCT was 555.15±29.648 μm. Clear positive correlations between GAT and I-Care, GAT and Tono-Pen, and I-Care and Tono-Pen tonometers were found (r=0.867, P0.05). Both non-gold standard tonometers were affected by CCT; that is, both I-Care and Tono-Pen tonometer values were significantly higher with higher CCT means (>555 μm; MD=−1.282, P555 μm) were associated with overestimated IOP values.