PLoS ONE (Jan 2014)

Biomechanical analysis of a newly developed shape memory alloy hook in a transforaminal lumbar interbody fusion (TLIF) in vitro model.

  • Xi Wang,
  • Jing Xu,
  • Yuexing Zhu,
  • Jiukun Li,
  • Si Zhou,
  • Shunliang Tian,
  • Yucheng Xiang,
  • Xingmo Liu,
  • Ying Zheng,
  • Tao Pan

DOI
https://doi.org/10.1371/journal.pone.0114326
Journal volume & issue
Vol. 9, no. 12
p. e114326

Abstract

Read online

The objective of this biomechanical study was to evaluate the stability provided by a newly developed shape memory alloy hook (SMAH) in a cadaveric transforaminal lumbar interbody fusion (TLIF) model.Six human cadaveric spines (L1-S2) were tested in an in vitro flexibility experiment by applying pure moments of ±8 Nm in flexion/extension, left/right lateral bending, and left/right axial rotation. After intact testing, a TLIF was performed at L4-5. Each specimen was tested for the following constructs: unilateral SMAH (USMAH); bilateral SMAH (BSMAH); unilateral pedicle screws and rods (UPS); and bilateral pedicle screws and rods (BPS). The L3-L4, L4-L5, and L5-S1 range of motion (ROM) were recorded by a Motion Analysis System.Compared to the other constructs, the BPS provided the most stability. The UPS significantly reduced the ROM in extension/flexion and lateral bending; the BSMAH significantly reduced the ROM in extension/flexion, lateral bending, and axial rotation; and the USMAH significantly reduced the ROM in flexion and left lateral bending compared with the intact spine (p0.05). Stability provided by the USMAH compared with the UPS was not significantly different. ROMs of adjacent segments increased in all fixed constructs (p>0.05).Bilateral SMAH fixation can achieve immediate stability after L4-5 TLIF in vitro. Further studies are required to determine whether the SMAH can achieve fusion in vivo and alleviate adjacent segment degeneration.