JA Clinical Reports (Jan 2016)

Reiterative ventricular fibrillation caused by R-on-T during temporary epicardial pacing: a case report

  • Yuki Nakamori,
  • Takuma Maeda,
  • Yoshihiko Ohnishi

DOI
https://doi.org/10.1186/s40981-016-0029-6
Journal volume & issue
Vol. 2, no. 1
pp. 1 – 6

Abstract

Read online

Abstract Epicardial pacemaker wire insertion is standard following cardiothoracic surgery. However, undersensing of pacing wires may cause the R-on-T phenomenon, which induces ventricular fibrillation. We report a case of a male patient with severe mitral regurgitation scheduled for mitral valve replacement who experienced two ventricular fibrillation episodes related to the R-on-T phenomenon caused by undersensing of the epicardial pacing wire. Both undersensing events happened despite an appropriately low sensing threshold. Notably, the stimulated T wave followed the QRS of the premature ventricular contraction (PVC). This case suggests that a PVC’s R wave may be undersensed despite a low sensing threshold. This critical complication may have occurred because pacemakers sense R waves using a slew rate, which is the quotient of voltage over time. As a result, pacemakers may undersense wide QRS waves such as PVCs. Avoiding this dangerous phenomenon completely is not possible using epicardial pacemakers; therefore we recommend carefully adapting epicardial pacing especially when PVC waves occur frequently.