Annals of Pediatric Cardiology (Jan 2021)

Delivery sheath tip invagination – An unusual cause of failure to retrieve duct occluder

  • Anunay Gupta,
  • Saurabh Kumar Gupta,
  • Sivasubramanian Ramakrishnan,
  • Shyam S Kothari

DOI
https://doi.org/10.4103/apc.APC_165_20
Journal volume & issue
Vol. 14, no. 2
pp. 242 – 243

Abstract

Read online

Transcatheter closure of patent ductus arteriosus is the standard of care. Retrieval of a duct occluder device is generally easy until it is detached from the delivery cable. We report two instances of failed retrieval of the device due to sheath tip invagination. The report highlights the importance of prompt identification of the mechanism of unforeseen complications in managing it effectively.