Current Directions in Biomedical Engineering (Sep 2017)

In-ear photoplethysmography for central pulse waveform analysis in non-invasive hemodynamic monitoring

  • Tigges Timo,
  • Rockstroh Jonas,
  • Pielmuş Alexandru,
  • Klum Michael,
  • Feldheiser Aarne,
  • Hunsicker Oliver,
  • Orglmeister Reinhold

DOI
https://doi.org/10.1515/cdbme-2017-0122
Journal volume & issue
Vol. 3, no. 2
pp. 587 – 590

Abstract

Read online

In recent years, the analysis of the photoplethys-mographic (PPG) pulse waveforms has attracted much research focus. However, the considered signals are primarily recorded at the fingertips, which suffer from reduced peripheral perfusion in situations like hypovolemia or sepsis, rendering waveform analysis infeasible. The ear canal is not affected by cardiovascular centralization and could thus prove to be an ideal alternate measurement site for pulse waveform analysis. Therefore, we developed a novel system that allows for highly accurate photoplethysmographic measurements in the ear canal. We conducted a measurement study in order to assess the signal-to-noise ratio of our developed system Hereby, we achieved a mean SNR of 40.65 dB. Hence, we could show that our system allows for highly accurate PPG recordings in the ear canal facilitating sophisticated pulse waveform analysis. Furthermore, we demonstrated that the pulse decomposition analysis is also applicable to in-ear PPG recordings.