Journal of Imaging (Dec 2020)

Comparison of Thermal Neutron and Hard X-ray Dark-Field Tomography

  • Alex Gustschin,
  • Tobias Neuwirth,
  • Alexander Backs,
  • Manuel Viermetz,
  • Nikolai Gustschin,
  • Michael Schulz,
  • Franz Pfeiffer

DOI
https://doi.org/10.3390/jimaging7010001
Journal volume & issue
Vol. 7, no. 1
p. 1

Abstract

Read online

High visibility (0.56) neutron-based multi-modal imaging with a Talbot–Lau interferometer at a wavelength of 1.6 Å is reported. A tomography scan of a strongly absorbing quartz geode sample was performed with both the neutron and an X-ray grating interferometer (70 kVp) for a quantitative comparison. Small scattering structures embedded in the absorbing silica matrix were well resolved in neutron dark-field CT slices with a spatial resolution of about 300 μm. Beneficial effects, such as monochromaticity and stronger penetration power of the used neutron radiation, helped to avoid the beam hardening-related artificial dark-field signal which was present in the X-ray data. Both dark-field modalities show mostly the same structures; however, some scattering features appear only in the neutron domain. Potential applications of combined X-ray and neutron multi-modal CT enabling one to probe both the nuclear and the electron density-related structural properties are discussed. strongly absorbing samples are now accessible for the dark-field modality by the use of thermal neutrons.

Keywords