Diabetology & Metabolic Syndrome (Aug 2018)

Practical guidance for treatment of patients with diabetes using flash glucose monitoring: a pilot study

  • Keiko Koide,
  • Koichiro Azuma,
  • Jun Nojima,
  • Kenichiro Kodama,
  • Yoshihito Atsumi

DOI
https://doi.org/10.1186/s13098-018-0364-z
Journal volume & issue
Vol. 10, no. 1
pp. 1 – 7

Abstract

Read online

Abstract Background Flash glucose monitoring (FGM) is a factory-calibrated, blood glucose measuring sensor system for patients with diabetes. We aimed to investigate the correlation between the sensor glucose (SG) value obtained using an FGM device and the traditional self-monitoring of blood glucose (SMBG) value. Methods In 30 patients with diabetes under insulin treatment, SG and SMBG values were measured for 2 weeks, and the correlation between the values was analyzed. Results The mean number of accumulated measurements of SG values was 1223.2 ± 193.0, whereas that of the SMBG values was 49.2 ± 21.3. Although SG and SMBG values showed a favorable correlation (R 2 = 0.8413), SG values were lower than SMBG values by an average of 7.9 ± 29.8 mg/dL. The correlation patterns fell into four types: low type (SG values lower than SMBG values; n = 12), high type (SG values higher than SMBG values; n = 3), cross type (the slope of the two regression lines crossed at a certain measurement value; n = 14), and matching type (the values overlapped; n = 1). Conclusions Recognition of the characteristic correlation patterns between SG and SMBG values is indispensable for certified diabetes educators to provide appropriate treatment guidance to patients with diabetes.