Analytica (Apr 2024)
Headspace-Selected Ion Flow Tube Mass Spectrometry Workflows for Rapid Screening and Quantitation of Hazardous Volatile Impurities in Personal Care Products
Abstract
Personal care products (PCPs) are intended for regular application by consumers and therefore assuring the safety of these products is very important. Recently, benzene contamination has been highlighted in certain PCPs. The present study applies selected ion flow tube mass spectrometry (SIFT-MS) to a simultaneous headspace analysis of benzene, 1,4-dioxane, and formaldehyde—all known or suspected carcinogens—in nine haircare products with supporting qualitative analysis by gas chromatography–mass spectrometry (GC-MS). Headspace-SIFT-MS method development is compatible with the method of standard additions, which is necessary for the quantitation of volatile impurities in these complex emulsions. Benzene was quantified above the low-ng g−1 limit of quantitation (LOQ) in three products, dioxane above the sub-μg g−1 LOQ in all products, and formaldehyde above the low-μg g−1 LOQ in two products, providing a quantitative analysis at concentrations relevant to consumer safety. This study facilitated the development of generic workflows for SIFT-MS method development and application in routine analysis of PCPs. The assessment of workflows for SIFT-MS compared to a conventional GC-MS analysis suggests that 8- to 30-fold throughput enhancements may be possible for quantitative and screening analysis using SIFT-MS.
Keywords