Six Sigma Metrics: An Evolving Indicator of Quality Assurance for Clinical Biochemistry

  • Devi Prasad Pradhan,
  • Debasish Pandit,
  • Sibasish Sahoo,
  • Roma Rattan,
  • Sucharita Mohanty

DOI
https://doi.org/10.7860/JCDR/2022/56691.16718
Journal volume & issue
Vol. 16, no. 8
pp. BC14 – BC18

Abstract

Read online

Introduction: The analytical phase of the total testing process is the one in which the clinical biochemist can directly intervene to improve the quality of tests reporting. The sigma metrics and Operational Process Specification (OPSpec) chart can specify to which category the laboratory belongs. Aim: To apply sigma metrics to analytical process of testing, do the root cause analysis and apply the corrective measures according to Westgard rules to improve laboratory performance towards the quality assurances. Materials and Methods: This was a retrospective-prospective study carried out in a clinical laboratory of MKCG Medical College and Hospital, Berhampur, Odisha, India, from July 2020 to March 2021. A retrospective secondary data analysis of six months duration was carried out in a clinical chemistry laboratory with a follow-up prospective study for three months. During this period, 16 analytes were tabulated to analyse the Internal Quality Control (IQC). External Quality Control (EQC) for the same analytes were obtained on monthly basis and the sigma metrics was calculated for each analytes. For analytes with sigma value <3, appropriate measures were taken according to Westgard rules to improvise the quality of laboratory investigations. The statistical analysis of sigma metrics was performed in “R” v-3.6.3. Results: Out of total 16 analytes, three analytes at level 1 and two analytes at level 2 Quality control (QC) showed a world class performance whereas four analytes showed a poor performance at both the QC levels with sigma metrics value <3. From Quality Goal Index (QGI) and root cause analysis, the source of error was detected and corrected. Conclusion: The inaccuracy and imprecision of different parameters in the analytical phase of the testing process can be addressed by calculating the sigma metrics and do the root cause analysis. Application of corrective measures according to Westgard rule can improve the laboratory performance towards the quality assurance.

Keywords