PLoS ONE (Jan 2017)

Minor defects of the luminal integrity in arterial introducer eSheaths after transcatheter aortic valve implantation.

  • Michael Buege,
  • Till Koehler,
  • Ralf Heiderhoff,
  • Marc Papenheim,
  • Si Wang,
  • Heinrich Schleiting,
  • Wolfgang H Arnold,
  • Jason R Foerst,
  • Melchior Seyfarth,
  • Klaus Tiroch,
  • Thomas Riedl,
  • Marc Vorpahl

DOI
https://doi.org/10.1371/journal.pone.0176893
Journal volume & issue
Vol. 12, no. 5
p. e0176893

Abstract

Read online

Medical devices such as implant delivery systems are commonly used during minimally invasive procedures in the cardiovascular system. These devices often have lubricious polymer coatings to reduce friction between the device and blood vessels but coatings may separate and potentially cause serious injuries to patients.Lubricious coated eSheaths for transcatheter heart valve implantation were assessed for luminal integrity at the proximal, medial and distal part. We assessed the number, depths and area of luminal trails using environmental scanning electron microscope (ESEM), white light interferometry (WLI) and optical profilometry using area scale fractal complexity (asfc) as surface parameters. A total of 15 eSheaths were retrieved and analyzed after successful femoral transcatheter Sapien 3 implantation in patients (23 mm valve- 14F eSheath, 26 mm valve- 14F eSheath and 29 mm valve- 16F eSheath, n = 5 for each group). Unused eSheaths (14F and 16F) served as controls (n = 5 for each group).ESEM revealed significantly greater number of trails after TAVR passage with the 23 mm, 26 mm and 29 mm valves compared to unused control 14F and 16F eSheaths (13.9 ± 3.1, 14.2 ± 2.3, 15.8 ± 1.7 vs. 0.08 ± 0.1 and 1.0 ± 0.5 [n]; p ≤ 0.0001 for all comparisons). Similarly, WLI showed minor, but significantly greater areas of luminal defects after 23 mm, 26 mm and 29 mm valve implantation vs. 14F and 16F unused controls (7.5 ± 0.9, 10.3 ± 1.1, 10.4 ± 1.4 vs. 4.1 ± 0.4 and 2.2 ± 0.4 [μm2], p = 0.0081). Likewise, the 3D-surface-measurement showed comparable results after implantation of the 23 mm, 26 mm and 29 mm valves vs. 14F and 16F unused control eSheaths (79.5 ± 6.3, 105.9 ± 5.3, 98.8 ± 4.8 vs. 5.1 ± 2.8 and 5.6 ± 0.5 [asfc] p = 0.0001).Measurable defects of the luminal layer occur during balloon expandable TAVR using 14F and 16F eSheaths though this is likely clinically insignificant. Further clinical investigations including a prospective assessment of minor peripheral embolization are needed to fully address the impact of this luminal defects.