Chinese Journal of Magnetic Resonance (Mar 2021)

A Design of Short Dead-Time RF Coil and RF Switch for Low-Field NMR

  • FENG Tao,
  • CHEN Jun-fei,
  • ZHANG Zhen,
  • YANG Chun-sheng,
  • ZHANG Zhi,
  • LIU Chao-yang

DOI
https://doi.org/10.11938/cjmr20202823
Journal volume & issue
Vol. 38, no. 01
pp. 1 – 11

Abstract

Read online

Based on the ring-down signal generation principle of nuclear magnetic resonance (NMR), we presented here a design scheme of transmitting and receiving separated short dead-time radio frequency (RF) coil composed of loop-gap resonator (LGR) coil and solenoid coil in low-field NMR system, to improve transmission efficiency by optimizing tuning and matching network. We also designed a fast RF switch and driver to fulfill the requirement of the RF coil. Based on the simulation results, the short dead-time RF coil was fabricated and applied to a self-developed 9.51 MHz portable low-field NMR spectrometer. The NMR experiments showed that the dead-time was reduced to less than 10 μs, which verified the feasibility of the design scheme.

Keywords