BIO Web of Conferences (Jan 2023)

Study on Calibration of Artificial Climate Chamber for Biological Application

  • Gong Wei,
  • Zhang Wen,
  • Ren Tingting,
  • Zeng Lingjian,
  • Feng Li

DOI
https://doi.org/10.1051/bioconf/20235903007
Journal volume & issue
Vol. 59
p. 03007

Abstract

Read online

Artificial climate chamber for biological application is widely used in biological laboratories. Parameters such as temperature, humidity and light intensity of artificial climate chamber can directly decide the accuracy of the experimental results. At present, there is no relevant calibration specification, or method for artificial climate chamber in China. This article presents a novel calibration method for artificial climate chambers for biological application to evaluate temperature, humidity, and light intensity control performance of the equipment, by determine corresponding environmental conditions, calibration standards, location of standards and calculation equations. The calibration results show that the calibration method is scientific and practical for metrological performance evaluation of artificial climate chambers for biological application.