Applied Sciences (Nov 2023)
The Development of an Interface Instrument for Collecting Electromyography Data and Controlling a Continuous Passive Motion Machine
Abstract
There is a lack of research in using electromyography (EMG) signals to control a continuous passive motion (CPM) machine. This study aimed to develop an interface instrument for digitalising EMG signals and controlling a CPM machine. Methods: The proposed device was designed with the following: (1) a signal processing unit which converted the EMGs from analogue to digital for the controller; (2) a personal computer which stored and displayed the EMG signals; (3) an LCD device to display the running angle of the CPM; and (4) a microcontroller unit to control the input/output signals and process the algorithm, driving the CPM. To validate the reliability of the proposed system, a total of 600 EMG trials were collected from 10 healthy subjects by using the proposed device via the Delsys® TringoTM EMG system and simultaneously using the Vicon® motion capture system. Result: This proposed device was able to digitalise and process EMG signals from eight channels of muscles, and the signals were able to drive a CPM. The validated results showed that the digitalised EMG signals by the proposed device were statistically similar to and correlated with the signals by the Vicon system with a median correlation coefficient of 0.81, with the 25% and 75% range being 0.56–0.92 with all pairs (300 pairs of EMG trials) (p < 0.001). Conclusions: This study confirmed that the developed device can digitalise EMG signals and drive a CPM as an applicable prototype that can work as an interface between EMG and CPM devices with high reliability.
Keywords