Separations (May 2024)

Coordination Ion Spray for Analysis of the Growth Hormones Releasing Peptides in Urine—An Application Study

  • Azamat Temerdashev,
  • Elina Gashimova,
  • Alice Azaryan,
  • Yu-Qi Feng,
  • Sanka N. Atapattu

DOI
https://doi.org/10.3390/separations11050155
Journal volume & issue
Vol. 11, no. 5
p. 155

Abstract

Read online

In this article, a comparison of ionization techniques is provided and discussed. Conventional liquid chromatography with an electrospray ionization source shows higher robustness and repeatability in comparison with liquid chromatography coupled with a coordination ion spray (CIS-MS) source using silver nitrate as the dopant. However, the higher sensitivity and possibility to collect more data in untargeted applications mean CIS-MS is emerging as an instrument used in specific applications. During this research, the limit of detection (LOD) for GHRP-2 and GHRP-6 was established at 0.2 ng/mL, and the lower limit of quantification (LLOQ) was 0.5 ng/mL for CIS-MS. For conventional ESI-MS combined with solid-phase extraction on weak cation exchange columns, the limit of detection was found to be 1 ng/mL, and the lower limit of quantification was 2 ng/mL.