Scientific Reports (Sep 2023)
HU-based material conversion for BNCT accurate dose estimation
Abstract
Abstract NeuMANTA is a new generation boron neutron capture therapy (BNCT)-specific treatment planning system developed by the Neuboron Medical Group and upgraded to an important feature, a Hounsfield unit (HU)-based material conversion algorithm. The range of HU values was refined to 96 specific groups and established corresponding to tissue information. The elemental compositions and mass densities have an important effect on the calculated dose distribution. The region of interest defined in the treatment plan can be converted into multiple material compositions based on HU values or assigned specified single material composition in NeuMANTA. Different material compositions may cause normal tissue maximum dose rates to differ by more than 10% in biologically equivalent doses and to differ by up to 6% in physically absorbed doses. Although the tumor has a lower proportion of BNCT background dose, the material composition difference may affect the minimum dose of biologically equivalent dose and physically absorbed dose by more than 3%. In addition, the difference in material composition could lead to a change in neutron moderation as well as scattering. Therefore, the material composition has a significant impact on the assessment of normal tissue side effects and tumor control probability. It is essential for accurate dose estimation in BNCT.