Physchem (Nov 2022)

The True Nature of the Energy Calibration for Nuclear Resonant Vibrational Spectroscopy: A Time-Based Conversion

  • Hongxin Wang,
  • Yoshitaka Yoda,
  • Jessie Wang

DOI
https://doi.org/10.3390/physchem2040027
Journal volume & issue
Vol. 2, no. 4
pp. 369 – 388

Abstract

Read online

Nuclear resonant vibrational spectroscopy (NRVS) is an excellent synchrotron-based vibrational spectroscopy. Its isotope specificity and other advantages are particularly good to study, for example, iron center(s) inside complicated molecules such as enzymes. In order to investigate some small energy shifts, the energy scale variation from scan to scan must be corrected via an in-situ measurement or with other internal reference peak(s) inside the spectra to be calibrated. On the other hand, the energy re-distribution within each scan also needs attention for a sectional scan which has a different scanning time per point in different sections and is often used to measure weak NRVS signals. In this publication, we: (1) evaluated the point-to-point energy re-distribution within each NRVS scan or within an averaged scan with a time-scaled (not energy-scaled) function; (2) discussed the errorbar contributed from the improper “distribution” of ΔEi or the averaged ΔE within one scan (Eerr1) vs. that due to the different ΔEi from different scans (Eerr2). It is well illustrated that the former (Eerr1) is as important as, or sometimes even more important than, the latter (Eerr2); and (3) provided a procedure to re-calibrate the published NRVS-derived PVDOS spectra in case of need. This article establishes the concept that, at least for sectional NRVS scans, the energy positions should be corrected according to the time scanned rather than be scaled with a universal constant, as in a conventional calibration procedure.

Keywords