Indian Pacing and Electrophysiology Journal (Mar 2022)

Intrinsic anti-tachycardia pacing terminated ventricular tachycardia resistant to traditional anti-tachycardia pacing

  • Hikaru Hagiwara,
  • Masaya Watanabe,
  • Rui Kamada,
  • Toshihisa Anzai

Journal volume & issue
Vol. 22, no. 2
pp. 99 – 102

Abstract

Read online

Implantable cardioverter-defibrillators (ICDs) serve to reduce the risk of sudden death; however, ICD shocks worsen patient prognosis. Therefore, attempts have been made to terminate life-threatening arrhythmias without ICD shocks. A 71-year-old man with non-ischemic cardiomyopathy, who previously underwent cardiac resynchronization therapy-defibrillator (CRT-D) placement, was hospitalized for ventricular tachyarrhythmia (VT) that was refractory to traditional anti-tachycardia pacing (ATP). Endocardial and epicardial ablation failed to prevent VT recurrence. Since the CRT-D battery was exhausted, it was replaced with a Cobalt™ XT HF CRT-D (Medtronic, Minneapolis, MN, USA), and the intrinsic ATP (iATP) algorithm was employed. Although VT recurred frequently, recurrent VTs were terminated by the iATP, which created a conduction block in the circuit without VT acceleration or shock. This is the first reported case wherein an iATP algorithm was effective against VT resistant to traditional anti-tachycardia pacing. This novel ATP algorithm has the potential to terminate refractory VT without ICD shocks and provide a better prognosis.

Keywords