Nuclear Technology and Radiation Protection (Jan 2008)

Modeling of germanium detector and its sourceless calibration

  • Steljić Milijana,
  • Milošević Miodrag,
  • Beličev Petar

DOI
https://doi.org/10.2298/NTRP0802051S
Journal volume & issue
Vol. 23, no. 2
pp. 51 – 57

Abstract

Read online

The paper describes the procedure of adapting a coaxial high-precision germanium detector to a device with numerical calibration. The procedure includes the determination of detector dimensions and establishing the corresponding model of the system. In order to achieve a successful calibration of the system without the usage of standard sources, Monte Carlo simulations were performed to determine its efficiency and pulse-height response function. A detailed Monte Carlo model was developed using the MCNP-5.0 code. The obtained results have indicated that this method represents a valuable tool for the quantitative uncertainty analysis of radiation spectrometers and gamma-ray detector calibration, thus minimizing the need for the deployment of radioactive sources.