Electrochemistry (Nov 2020)

Determination of Lactate by Sequential Injection Analysis Using a Fluoride Ion-selective Electrode Detector

  • Yuichi TAKAHASHI,
  • Takashi MASADOME

DOI
https://doi.org/10.5796/electrochemistry.20-00097
Journal volume & issue
Vol. 88, no. 6
pp. 522 – 524

Abstract

Read online

A combination of sequential Injection Analysis (SIA) system and the ion-selective electrode (ISE) detector is considered to be one of a very effective analytical method. Reaction of lactate with lactate oxidase produces pyruvic acid and hydrogen peroxide. When the produced hydrogen peroxide is reacted with 4-aminoantipyrine (4-AAP) and 4-fluorophenol (4-FP) in the presence of horseradish peroxidase (HRP), F− ion is produced. The concentration of lactate can be indirectly determined by measuring the F− ion by F− ISE. Based on this principle of detection of lactate, a new measurement method of lactate using an SIA with ISE detector was developed.

Keywords