International Journal of Infectious Diseases (Sep 2022)

Accuracy of QuantiFERON SARS-CoV-2 research use only assay and characterization of the CD4+ and CD8+ T cell-SARS-CoV-2 response: comparison with a homemade interferon-γ release assay

  • Alessandra Aiello,
  • Andrea Coppola,
  • Valentina Vanini,
  • Linda Petrone,
  • Gilda Cuzzi,
  • Andrea Salmi,
  • Anna Maria Gerarda Altera,
  • Carla Tortorella,
  • Gina Gualano,
  • Claudio Gasperini,
  • Palma Scolieri,
  • Alessia Beccacece,
  • Serena Vita,
  • Vincenzo Bruzzese,
  • Roberto Lorenzetti,
  • Fabrizio Palmieri,
  • Emanuele Nicastri,
  • Delia Goletti

Journal volume & issue
Vol. 122
pp. 841 – 849

Abstract

Read online

Objectives: In this study, we aimed to characterize the SARS-CoV-2-specific T cell response detected by the QuantiFERON SARS-CoV-2 research use only assay in terms of accuracy and T cell subsets involved compared with a homemade interferon (IFN)-γ release assay (IGRA).Methods: We evaluated T cell response by the standardized QuantiFERON SARS-CoV-2 tubes (antigen [Ag]1 and Ag2) and a homemade IGRA quantifying IFN-γ response to SARS-CoV-2 spike peptides (homemade-IGRA-SPIKE test). We evaluated the T cell subsets mediating the specific response using flow cytometry.Results: We prospectively enrolled 66 individuals: COVID-19 or post-COVID-19 subjects and NO-COVID-19-vaccinated subjects, including healthy donors and immunocompromised subjects. The standardized kit detected 62.1% (41/66) of T cell responders. Ag2 tube showed a higher IFN-γ quantitative and qualitative response. Ag1 tube response was mainly mediated by CD4+ T cells; Ag2 tube response was mediated by CD4+ and CD8+ T cells. The homemade-IGRA-SPIKE test detected a higher number of responders (52/66, 78.8%) than the QuantiFERON SARS-CoV-2 assay (P = 0.056). The response was found in both T cell subsets, although a higher magnitude and response rate was observed in the CD4+ T cell subset.Conclusion: The QuantiFERON SARS-CoV-2 response is mediated by CD4+ and CD8+ T cells. A lower number of responders is found compared with the homemade-IGRA-SPIKE test, likely because of the different peptide composition.

Keywords