Applied Sciences (Apr 2022)

Accuracy of Additively Manufactured Dental Casts Compared with That of Virtual Scan Data Obtained with an Intraoral Scanner: An In Vitro Study

  • Gi Youn Kim,
  • Jiayi Li,
  • Hong Seok Moon,
  • Kyung Chul Oh

DOI
https://doi.org/10.3390/app12094348
Journal volume & issue
Vol. 12, no. 9
p. 4348

Abstract

Read online

The study aimed to evaluate the time-related accuracy of additively manufactured dental casts and to compare it with scan data obtained with an intraoral scanner in vitro. Twenty-eight markers were attached to a set of dentiforms as reference model, and the distances between the markers were measured using a digital caliper. An intraoral scanner was used to obtain the virtual scan data of the reference model with a total of 30 scans per arch. The distances between markers were measured using a three-dimensional inspection software for all scans (group IOS). Scan data were additively manufactured using a 3D printer, and the distances between markers were measured as in the reference model immediately after post-polymerization (group PPIA), 1 day (group PP1D), 7 days (group PP7D), and 30 days after post-polymerization (group PP30D). The linear deviation in group IOS was 199.74 ± 11.14 μm, PPIA was 242.88 ± 49 μm, PP1D was 259.9 ± 42.59 μm, PP7D was 289.82 ± 39.74 μm, and PP30D was 315.8 ± 33.28 μm, in comparison with the reference model, with significant differences among all groups (all p < 0.05). When additively manufacturing casts from scan data to verify the quality of dental prostheses designed virtually, the prostheses should be adapted to casts manufactured within one week.

Keywords