Abstract

Read online

Intra Aortic Balloon Pump (IABP) is conventionally used to support coronary perfusion and weaning from cardiopulmonary bypass. IABP in situ has its own share of complications. We present a case where a patient on IABP support who had reduced peripheral pulsations of the ipsilateral limb and was initially misdiagnosed as IABP catheter associated thromboembolism. A negative embolectomy ruled out the same. Further looking for the cause of reduction of ipsilateral pulses it was found that the tight compressive bandage at saphenous vein conduit harvesting site had led to development of compartment syndrome (CS).