Determination of neutron spectrum outside the lateral shielding of Taiwan Photon Source using high-sensitivity Bonner cylinders

  • Lin Yu-Chi,
  • Chen Ang-Yu,
  • Liu Joseph,
  • Lee Kuo-Wei,
  • Sheu Rong-Jiun

DOI
https://doi.org/10.1051/epjconf/201715303012
Journal volume & issue
Vol. 153
p. 03012

Abstract

Read online

A homemade Bonner cylinder spectrometer was used to determine the neutron spectrum from thermal energy up to several GeV at a heavily-shielded light source facility. The spectrometer, similar to the design of Bonner spheres, features high sensitivity of neutron detection because of using a long cylindrical 3He proportional counter. The spectrum measurement was performed during the facility commissioning by intentionally parking the injected electrons at the septum of the storage ring. Based on a high-fidelity FLUKA simulation, the predicted neutron spectrum at the location of measurement was adopted as the initial guess of spectrum unfolding. The unfolded result indicated an underestimation of the calculated neutron spectrum in the high-energy portion, leading to a substantial revision of the neutron dose rate at the location.