Romanian Journal of Military Medicine (Jul 2016)

Finite element analysis of a modified short hip endoprosthesis

  • Augustin Semenescu,
  • Florentina Ioniță Radu,
  • Ileana M. Mateș,
  • Petre Bădică,
  • Nicolae D. Batalu

Journal volume & issue
Vol. CXIX, no. 2
pp. 27 – 31

Abstract

Read online

A finite element simulation of the mechanical static features for a modified short hip endoprosthesis was performed. The corkscrew-like femoral stem was modified introducing more turns of the thread. By such an approach it is expected that for some cases the mechanical fixation of the prosthesis to the bone will be improved or the use of the cement for bonding is not necessary. Our scenario was estimated for titanium and stainless steel, and both materials show good safety factors. Mechanical stress is expected to be distributed more uniform in the bone for the new design with more turns of thread.