P-shaped Coiled Stator Ultrasound Motor for Rotating Intravascular Surgery Device

  • Toshinobu ABE,
  • Tadashi MORIYA,
  • Takasuke IRIE,
  • Masakazu SATOU,
  • Shinichi TAKEUCHI

Journal volume & issue
Vol. 184, no. 1
pp. 108 – 115

Abstract

Read online

The primary focus of this paper is the development of an ultra-miniature ultrasound motor for use in the human blood vessel. Since the size of the drive source for rotating the atherectomy device and intravascular ultrasonography system are large currently in practical use, it is installed outside the body, and the rotational power for the atherectomy device and intravascular ultrasonography system are transmitted through the long tortuous blood vessel. Such systems suffer from the problem that the rotation becomes non-uniform, and the problem that the available time is limited. We have therefore developed a P-shaped coiled stator ultrasound motor as a miniature ultrasound motor for rotating the ultrasound sensor for use in blood vessels in order to solve these problems. In this paper, we describe measurement of the torque, revolution speed, output power, efficiency, and particle motion on acoustic waveguide of the P-shaped coiled stator ultrasound motor.