Molecules (Jun 2024)

Development of an RP-HPLC Method for Quantifying Diclofenac Diethylamine, Methyl Salicylate, and Capsaicin in Pharmaceutical Formulation and Skin Samples

  • Lubna H. M. AlNahwa,
  • Hazim M. Ali,
  • Tamer H. A. Hasanin,
  • Khaled Shalaby,
  • Mutairah S. Alshammari,
  • Alaa M. Alsirhani,
  • Sabrein H. Mohamed

DOI
https://doi.org/10.3390/molecules29122732
Journal volume & issue
Vol. 29, no. 12
p. 2732

Abstract

Read online

An RP-HPLC method with a UV detector was developed for the simultaneous quantification of diclofenac diethylamine, methyl salicylate, and capsaicin in a pharmaceutical formulation and rabbit skin samples. The separation was achieved using a Thermo Scientific ACCLAIMTM 120 C18 column (Waltham, MA, USA, 4.6 mm × 150 mm, 5 µm). The optimized elution phase consisted of deionized water adjusted to pH = 3 using phosphoric acid mixed with acetonitrile in a 35:65% (v/v) ratio with isocratic elution. The flow rate was set at 0.7 mL/min, and the detection was performed at 205 nm and 25 °C. The method exhibits good linearity for capsaicin (0.05–70.0 µg/mL), methyl salicylate (0.05–100.0 µg/mL), and diclofenac diethylamine (0.05–100.0 µg/mL), with low LOD values (0.0249, 0.0271, and 0.0038 for capsaicin, methyl salicylate, and diclofenac diethylamine, respectively). The RSD% values were below 3.0%, indicating good precision. The overall greenness score of the method was 0.61, reflecting its environmentally friendly nature. The developed RP-HPLC method was successfully applied to analyze Omni Hot Gel® pharmaceutical formulation and rabbit skin permeation samples.

Keywords