Micromachines (Mar 2022)

An Endovascular Catheterization Robotic System Using Collaborative Operation with Magnetically Controlled Haptic Force Feedback

  • Xinming Li,
  • Shuxiang Guo,
  • Peng Shi,
  • Xiaoliang Jin,
  • Masahiko Kawanishi

DOI
https://doi.org/10.3390/mi13040505
Journal volume & issue
Vol. 13, no. 4
p. 505

Abstract

Read online

Robot-assisted technology is often used to perform endovascular catheterization surgeries, which generally depend on the flexible operability and the accurate force feedback of a robotic system. In this paper, an endovascular catheterization robotic system (ECRS) was developed to improve collaborative operation and haptic force feedback. A couple of operating handles were designed to maximize the use of the natural operations of surgeons on the master side, which is a flexible and ergonomic device. A magnetically controlled haptic force feedback structure is proposed based on hydrogel and solid magnetorheological (MR) fluid to offer a sense of haptic feedback to operators; this has potential influence on the field of force feedback. In addition, a unique tremor-reduction structure is introduced to enhance operating safety. Tracking performance experiments and in vitro experiments were conducted to evaluate the performance of the developed ECRS. According to these experimental results, the average translation-tracking error is 0.94 mm, and the average error of rotation is 0.89 degrees. Moreover, in vitro experiments demonstrated that haptic feedback has the advantage of reducing workload and shortening surgery completion time. The developed ECRS also has the benefits of inspiring other researchers to study collaborative robots and magnetically controlled feedback.

Keywords