Journal of Interventional Cardiology (Jan 2022)

Balloon-Expandable TAVR Bioprostheses: Area or Perimeter Sizing? A Prospective Pilot Study

  • Jonathan Halim,
  • Peter den Heijer,
  • Jeroen Vos,
  • Bas E. Schölzel,
  • Martijn Meuwissen,
  • Ben van den Branden,
  • Andreas Baumbach,
  • Alexander J. J. Ijsselmuiden

DOI
https://doi.org/10.1155/2022/3139476
Journal volume & issue
Vol. 2022

Abstract

Read online

Objective. In TAVR, area sizing is used for balloon-expandable (BE) valves, whereas self-expanding valves are sized to annulus perimeter. For BE valves, this seems illogical: these frames force a circular shape even on an ellipsoid annulus. This can potentially lead to relative undersizing when area sizing is being applied. We developed a perimeter-based sizing algorithm to evaluate the safety and feasibility of perimeter sizing for the Myval BE valve. Methods. In this prospective single-center study, 60 patients with severe aortic stenosis treated with the Myval BE valve were included. Perimeter sizing was used with limited oversizing of 3.7% ± 1.3% compared to the annulus perimeter. After TAVR, clinical outcomes were evaluated at 30 days and 1 year. An echocardiographic follow-up took place at 30 days. Results. At 30 days, the need for PPI and stroke occurred in 2% and 3% of the patients, respectively. Moreover, cardiac death and moderate-severe PVL were absent. At 1-year, cardiac death and stroke were observed in 3% and 8% of the patients, respectively. In 33.3% of the patients, a larger valve size was implanted compared to the valve size calculated by area sizing. Conclusions. Perimeter sizing with the Myval BE valve leads to substantial use of larger valve sizes and favorable clinical outcomes, with low PPI and the absence of significant PVL. A randomized controlled trial is being planned to prove the superiority of this alternative sizing method.