Renal Replacement Therapy (Oct 2022)

High-volume pre-dilution on-line hemodiafiltration is the adequate blood purification method from the viewpoint of amino acid nutrition

  • Motoko Kato,
  • Shunichiro Urabe,
  • Shohei Matsuzawa,
  • Akira Kato,
  • Momoko Fukazawa,
  • Emi Hiyama,
  • Asami Kurii,
  • Noriko Mikami,
  • Yukie Kitajima,
  • Miho Hida,
  • Kenichi Kokubo,
  • Toru Hyodo

DOI
https://doi.org/10.1186/s41100-022-00441-4
Journal volume & issue
Vol. 8, no. 1
pp. 1 – 5

Abstract

Read online

Abstract Background We compared amino acid leakage among high-volume pre-dilution on-line hemodiafiltration (Pre O-HDF), hemodialysis (HD), and post-dilution on-line hemodiafiltration (Post O-HDF). Subjects and methods The subjects were 9 patients. For HD, the total dialysate flow rate was established as 500 mL/min. For high-volume Pre O-HDF, it was established as 600 mL/min, under a replacement fluid volume of 90 L. For Post O-HDF, it was established as 600 mL/min, under a replacement fluid volume of 10 L. In both procedures, the duration of treatment was 4 h, and the blood flow volume was 200 mL/min. We compared the leakages of total amino acid, non-essential amino acid, and essential amino acid, clear spaces (CSs), β2-microglobulin (β2-MG) reduction rate, Kt/V for urea, and albumin leakage among the three procedures. Results Amino acid leakages after high-volume Pre O-HDF were significantly lower than HD and Post O-HDF. The CSs after high-volume Pre O-HDF were significantly lower than Post O-HDF. The β2-MG reduction rate after high-volume Pre O-HDF was significantly lower than Post O-HDF. The Kt/V for urea was not significantly different. Albumin leakages were below the detection limit (< 1 g) in the three procedures. Conclusion Under the treatment conditions we performed this time, high-volume Pre O-HDF reduces amino acid leakage in comparison with HD and Post O-HDF. High-volume Pre O-HDF is a therapeutic mode that suppresses amino acid leakage.

Keywords