SAGE Open Medical Case Reports (Sep 2016)

Delayed Type IIIb endoleak secondary to graft fabric tear 7 years following implantation of a Medtronic Talent endovascular aortic device: A case report and review of the literature

  • Vinay Kansal,
  • Sudhir Nagpal

DOI
https://doi.org/10.1177/2050313X16670304
Journal volume & issue
Vol. 4

Abstract

Read online

Objectives: To report a rare case of delayed Type IIIb endoleak secondary to fabric tear following implantation of a Medtronic Talent endovascular device. Methods: A 83-year old gentleman underwent elective endovascular aneurysm repair for infrarenal abdominal aortic aneurysm with a Medtronic bifurcated stent graft in 2008. Results: Seven years after the initial repair, imaging surveillance revealed significant endoleak and brisk aneurysm sac expansion due to Type IIIb endoleak secondary to endograft limb fabric tear. Conclusions: This case illustrates the imperative role of imaging surveillance in detection of long-term endovascular aneurysm repair complications. Furthermore, we discuss exclusion of the graft tear with aortouniiliac stent grafting as the treatment for this complication.