Journal of Marine Science and Engineering (Mar 2024)

A New Method for the Correction of Spectrum Drift Caused by Temperature Changes When Using a NaI(Tl) Seawater Radioactivity Sensor

  • Yan Shi,
  • Yingying Zhang,
  • Xiandong Feng,
  • Da Yuan,
  • Bingwei Wu,
  • Haijie Bi

DOI
https://doi.org/10.3390/jmse12040546
Journal volume & issue
Vol. 12, no. 4
p. 546

Abstract

Read online

The seawater radioactivity sensor based on the NaI(Tl) scintillation crystal detection method is prone to spectrum drift due to environment temperature changes during actual measurement, which pose difficulties in the activity analysis of seawater radionuclides. This paper adopts experiment design and analysis methods to study the relationship laws between temperature and peak position channel address and between gain and the peak position channel address. On the basis of studying and summarizing the two relationship laws of the NaI(Tl) seawater radioactivity sensor, a spectrum drift correction method combining gain adjustment and spectrum processing is proposed. Laboratory air and water environment experiments verify that the peak position channel address drift after air environment correction does not exceed ±2 channels, and after water environment correction does not exceed ±1 channel. Long-term verification experiments at the seawater work site have shown that this correction method can effectively correct the spectrum drift of the NaI(Tl) seawater radioactivity sensor and meet the requirements of long-term automatic operations at sea for the measured spectrum.