Integration of a Miniature Quartz Crystal Microbalance with a Microfluidic Chip for Amyloid Beta-Aβ42 Quantitation

  • Wenyan Tao,
  • Qingji Xie,
  • Hairui Wang,
  • Shanming Ke,
  • Peng Lin,
  • Xierong Zeng

DOI
https://doi.org/10.3390/s151025746
Journal volume & issue
Vol. 15, no. 10
pp. 25746 – 25760

Abstract

Read online

A miniature quartz crystal microbalance (mQCM) was integrated with a polydimethylsiloxane (PDMS) microfluidic device for on-chip determination of amyloid polypeptide–Aβ42. The integration techniques included photolithography and plasma coupling. Aβ42 antibody was immobilized on the mQCM surface using a cross-linker method, and the resonance frequency of mQCM shifted negatively due to antibody-antigen binding. A linear range from 0.1 µM to 3.2 µM was achieved. By using matrix elimination buffer, i.e., matrix phosphate buffer containing 500 µg/mL dextran and 0.5% Tween 20, Aβ42 could be successfully detected in the presence of 75% human serum. Additionally, high temperature treatments at 150 °C provided a valid method to recover mQCM, and PDMS-mQCM microfluidic device could be reused to some extent. Since the detectable Aβ42 concentration could be as low as 0.1 µM, which is close to cut-off value for Alzheimer patients, the PDMS-mQCM device could be applied in early Alzheimer’s disease diagnosis.