Journal of Functional Biomaterials (Jun 2024)
Enhanced Antibacterial Properties of Titanium Surfaces through Diversified Ion Plating with Silver Atom Deposition
Abstract
In this study, we investigate the antibacterial effect of silver atoms implanted into a thin surface layer of titanium at low energies using an alternative ion plating technology called Diversified Ion Plating. Silver atoms were incorporated into titanium samples using reactive low-voltage ion plating at 2 keV and 4 keV. Surface modifications and morphology were evaluated using wettability, profilometry measurements, and energy-dispersive spectroscopy. For a precise determination of the quantity and depth of implanted silver atoms on titanium surfaces, a combination of experimental techniques such as Rutherford Backscattering Spectrometry along with Monte Carlo simulations were utilized. To assess the antibacterial effects of the silver atoms incorporated into pure titanium surfaces, bacterial suspension immersion tests were performed with a standard strain of Staphylococcus aureus (ATCC 12600). The outcomes indicate that titanium surfaces implanted with silver atoms were more effective in inhibiting the growth of Staphylococcus aureus than pure titanium surfaces. Better results were found when the deposition was performed at 4 keV, indicating that a deeper implantation of silver, spanning a few nanometers, can result in a longer and more effective release of silver atoms. These findings suggest the potential for the development of new, cost-effective biomaterials, paving the way for improved implant materials in various health-related applications.
Keywords