Microbubbles as a contrast agent in grating interferometry mammography: an ex vivo proof-of-mechanism study

  • Kristina Lång,
  • Carolina Arboleda,
  • Serafino Forte,
  • Zhentian Wang,
  • Sven Prevrhal,
  • Thomas Koehler,
  • Norbert Kuhn,
  • Bernd David,
  • Konstantins Jefimovs,
  • Rahel A. Kubik-Huch,
  • Marco Stampanoni

DOI
https://doi.org/10.1186/s41747-019-0097-2
Journal volume & issue
Vol. 3, no. 1
pp. 1 – 6

Abstract

Read online

Abstract Grating interferometry mammography (GIM) is an experimental breast imaging method at the edge of being clinically implemented. Besides attenuation, GIM can measure the refraction and scattering of x-rays resulting in differential phase contrast (DPC) and dark-field (DF) images. In this exploratory study, we assessed the feasibility of using microbubbles as a contrast agent in GIM. Two millilitres of microbubbles and iodine were respectively injected into ex vivo breast phantoms, consisting of fresh chicken breasts. Native and postcontrast images were acquired with a clinically compatible GIM setup, operated at 38 kVp, 14-s acquisition time, and with a dose of 1.3 mGy. The visibility of the contrast agents was analysed in a side-by-side comparison by three radiologists. The contrast-to-noise-ratio (CNR) was calculated for each contrast agent. We found that both contrast agents were judged to be visible by the readers. The mean CNR was 3.1 ± 1.9 for microbubbles in DF and 24.2 ± 6.5 for iodine in attenuation. In conclusion, this is a first proof-of-mechanism study that microbubbles could be used as a contrast agent in clinically compatible GIM, due to their scattering properties, which implies the potential use of a contrast agent with a high safety profile in x-ray-based breast imaging.

Keywords