Journal of Cardiovascular Development and Disease (Apr 2024)
Risk of Cardiac Implantable Electronic Device Infection after Early versus Delayed Lead Repositioning
Abstract
(1) Background: Early reintervention increases the risk of infection of cardiac implantable electronic devices (CIEDs). Some operators therefore delay lead repositioning in the case of dislocation by weeks; however, there is no evidence to support this practice. The aim of our study was to evaluate the impact of the timing of reoperation on infection risk. (2) Methods: The data from consecutive patients undergoing lead repositioning in two European referral centers were retrospectively analyzed. The odds ratio (OR) of CIED infection in the first year was compared among patients undergoing early (≤1 week) vs. delayed (>1 week to 1 year) reoperation. (3) Results: Out of 249 patients requiring CIED reintervention, 85 patients (34%) underwent an early (median 2 days) and 164 (66%) underwent a delayed lead revision (median 53 days). A total of nine (3.6%) wound/device infections were identified. The risk of infection was numerically lower in the early (1.2%) vs. delayed (4.9%) intervention group yielding no statistically significant difference, even after adjustment for typical risk factors for CIED infection (adjusted OR = 0.264, 95% CI 0.032–2.179, p = 0.216). System explantation/extraction was necessary in seven cases, all being revised in the delayed group. (4) Conclusions: In this bicentric, international study, delayed lead repositioning did not reduce the risk of CIED infection.