International Journal of Nanomedicine (Aug 2015)

Measurement of creatinine in human plasma using a functional porous polymer structure sensing motif

  • Nanda SS,
  • An SSA,
  • Yi DK

Journal volume & issue
Vol. 2015, no. Special Issue on diverse applications in Nano-Theranostics
pp. 93 – 99

Abstract

Read online

Sitansu Sekhar Nanda,1 Seong Soo A An,1 Dong Kee Yi2,3 1Department of Bionanotechnology, Gachon Medical Research Institute, Gachon University, Seongnam, 2Department of Chemistry, 3Department of Environmental Engineering, Myongji University, Yongin, South Korea Abstract: In this study, a new method for detecting creatinine was developed. This novel sensor comprised of two ionic liquids, poly-lactic-co-glycolic acid (PLGA) and 1-butyl-3-methylimidazolium (BMIM) chloride, in the presence of 2',7'-dichlorofluorescein diacetate (DCFH-DA). PLGA and BMIM chloride formed a functional porous polymer structure (FPPS)-like structure. Creatinine within the FPPS rapidly hydrolyzed and released OH-, which in turn converted DCFH-DA to DCFH, developing an intense green color or green fluorescence. The conversion of DCFH to DCF+ resulted in swelling of FPPS and increased solubility. This DCF+-based sensor could detect creatinine levels with detection limit of 5 µM and also measure the creatinine in blood. This novel method could be used in diagnostic applications for monitoring individuals with renal dysfunction. Keywords: PLGA, FPPS, creatinine, blood, diagnosis