PLoS ONE (Jan 2016)

Direct Effective Dose Calculations in Pediatric Fluoroscopy-Guided Abdominal Interventions with Rando-Alderson Phantoms - Optimization of Preset Parameter Settings.

  • Moritz Wildgruber,
  • René Müller-Wille,
  • Holger Goessmann,
  • Wibke Uller,
  • Walter A Wohlgemuth

DOI
https://doi.org/10.1371/journal.pone.0161806
Journal volume & issue
Vol. 11, no. 8
p. e0161806

Abstract

Read online

The aim of the study was to calculate the effective dose during fluoroscopy-guided pediatric interventional procedures of the liver in a phantom model before and after adjustment of preset parameters.Organ doses were measured in three anthropomorphic Rando-Alderson phantoms representing children at various age and body weight (newborn 3.5kg, toddler 10kg, child 19kg). Collimation was performed focusing on the upper abdomen representing mock interventional radiology procedures such as percutaneous transhepatic cholangiography and drainage placement (PTCD). Fluoroscopy and digital subtraction angiography (DSA) acquisitions were performed in a posterior-anterior geometry using a state of the art flat-panel detector. Effective dose was directly measured from multiple incorporated thermoluminescent dosimeters (TLDs) using two different parameter settings.Effective dose values for each pediatric phantom were below 0.1mSv per minute fluoroscopy, and below 1mSv for a 1 minute DSA acquisition with a frame rate of 2 f/s. Lowering the values for the detector entrance dose enabled a reduction of the applied effective dose from 12 to 27% for fluoroscopy and 22 to 63% for DSA acquisitions. Similarly, organ doses of radiosensitive organs could be reduced by over 50%, especially when close to the primary x-ray beam.Modification of preset parameter settings enabled to decrease the effective dose for pediatric interventional procedures, as determined by effective dose calculations using dedicated pediatric Rando-Alderson phantoms.