Scientific Reports (Mar 2025)

Adaptive calibration of Gafchromic EBT3 film using generalized additive neural networks

  • Liyun Chang,
  • Pang-Yu Chen,
  • Shyh-An Yeh,
  • Chen-Lin Kang,
  • Chun-Ting Su,
  • Tsair-Fwu Lee

DOI
https://doi.org/10.1038/s41598-025-92568-7
Journal volume & issue
Vol. 15, no. 1
pp. 1 – 11

Abstract

Read online

Abstract In radiation therapy, precise dose distribution is essential for minimizing damage to normal tissues. Gafchromic EBT3 film is widely used to assure the quality of two-dimensional dosimetry but requires frequent recalibrations due to changes in sensitivity over time. This study presents a new calibration method using a Keras-based generalized additive neural network (GANN) to address film aging. EBT3 films from four lots were calibrated with a 6 MV photon beam and scanned on Epson scanners. The GANN method achieved percentage differences between calibrated and delivered doses within 5%, comparable to traditional recalibration methods, with an overall uncertainty of approximately 2%. It demonstrated improved stability and accuracy, significantly reducing the need for frequent recalibrations and providing a robust solution for long-term film dosimetry.