Scientific Reports (Jun 2024)

Comparison of the diagnostic accuracy of the Pluslife Mini Dock RHAM technology with Abbott ID Now and Cepheid GenXpert: A retrospective evaluation study

  • Laura Herrmann,
  • Juliana Breuer,
  • Tuan Ngo Duc,
  • Nicole Thomé,
  • Fatemeh Ghazaani,
  • Sundrela Kamhieh-Milz,
  • Julian Kamhieh-Milz,
  • Andreas Pfützner

DOI
https://doi.org/10.1038/s41598-024-64406-9
Journal volume & issue
Vol. 14, no. 1
pp. 1 – 9

Abstract

Read online

Abstract Rapid and sensitive detection of pathogens is critical in interrupting the transmission chain of infectious diseases. Currently, real-time (RT-)PCR represents the gold standard for the detection of SARS-CoV-2. RNase HII-assisted amplification (RHAM) is a promising technology, enabling reliable point-of-care (PoC) testing; however, its diagnostic accuracy has not yet been investigated. The present study compared the Pluslife Mini Dock (RHAM technology), with Abbott ID Now and Cepheid GeneXpert IV. The positive percent agreement (PPA) and negative percent agreement (NPA) were determined in 100 SARS-CoV-2 positive and 210 SARS-CoV-2 negative samples. Further, the reliability of the Pluslife Mini Dock was investigated in different SARS-CoV-2 variants (Delta and Omicron subvariants). The PPA was 99.00% for Pluslife, 100.00% for Abbott ID Now, and 99.00% for Cepheid GeneXpert, with an NPA of 100.00%, 98.90%, and 93.72%, respectively. Abbott ID Now demonstrated the highest rate of invalid results. All SARS-CoV-2 analysed variants were detected by the Pluslife device. Altogether, the Pluslife Mini Dock demonstrated a PPA of 99.16% (235/237) for CT < 36 and an NPA of 100.00% (313/313), respectively. In conclusion, the Pluslife Mini Dock demonstrated better analytical performance than Abbott ID Now and Cepheid GeneXpert IV, representing a highly accurate and rapid PoC alternative to RT-PCR.