Laryngoscope Investigative Otolaryngology (Aug 2023)

Endotracheal tube forces exerted on the larynx and a novel support device to reduce it

  • Andrew M. Vahabzadeh‐Hagh,
  • Brennan P. Marsh‐Armstrong,
  • Shiv H. Patel,
  • Luke Lindenmuth,
  • Zeyu Feng,
  • Rufu Gong,
  • Yun‐An Lin,
  • Taylor Pierce,
  • Kenneth J. Loh

DOI
https://doi.org/10.1002/lio2.1118
Journal volume & issue
Vol. 8, no. 4
pp. 989 – 995

Abstract

Read online

Abstract Objective Endotracheal tubes (ETTs) are commonly associated with laryngeal injury that may be short lasting and temporary or more severe and life altering. Injury is believed to result from forces that these ETTs exert on the larynx. Here we quantify the forces of ETTs of various sizes on the laryngotracheal complex to gain a more quantitative understanding of these potential damaging forces. Here we also perform preclinical testing of a novel support device to offload these forces. Methods Endotracheal intubation was performed on a fresh human cadaver using various ETT sizes. A strain‐sensitive graphene nanosheet sensor and a commercially available force sensing resistor were secured behind the larynx, anterior to the prevertebral fascia. The forces exerted on the larynx were measured for each of the commonly used ETTs. A novel support device, ETT clip (Endo Clip), was attached to the ETTs and changes in these forces were observed. Results Forces exerted on the laryngotracheal complex by various ETTs were observed to increase with increasing tube size. This pressure can be significantly reduced with a novel ETT clip. Conclusion Here we demonstrate the first quantitative measurement of forces that ETTs exert on the larynx. We demonstrate a novel device that can easily clip onto an ETT reducing pressure on the laryngotracheal complex. This preclinical test paves the way for a human clinical trial. Level of evidence 5.

Keywords