Journal of Orthopaedic Surgery and Research (Jan 2019)
Inferomedial cortical bone contact and fixation with calcar screws on the dynamic and static mechanical stability of proximal humerus fractures
Abstract
Abstract Background This study aimed to explore the effect of retaining inferomedial cortical bone contact and fixation with calcar screws on the dynamic and static mechanical stability of proximal humerus fractures treated with a locking plate. Methods Twelve Synbone prosthetic humeri (SYNBONE-AG, Switzerland) were used for a wedge osteotomy model at the proximal humerus, in four groups. In the cortex contact + screw fixation group and cortex contact group, the inferomedial cortical bone contact was retained. In the screw fixation group and control group, the inferomedial cortical bone contact was not retained. Calcar screw fixation was implemented only in the screw fixation groups. The dynamic and static mechanical stability of the models were tested with dynamic fatigue mechanics testing, quasi-static axial compression, three-point bending, and torsion testing. Results The cortex contact + screw fixation group showed the longest fatigue life and the best stability. There was 35% difference in fatigue life between the cortex contact + screw fixation group and the cortex contact group, 43%between the cortex contact + screw fixation group and screw fixation group, and 63% between the cortex contact + screw fixation group and screw fixation group (P < 0.01). The cortex contact + screw fixation group showed the best axial compressive stiffness, bending stiffness, and torsion stiffness; these were successively decreased in the other three groups (P < 0.01). Conclusion Retaining inferomedial cortical bone contact and fixation with two calcar screws maintained fracture stability with the highest strength and minimum deformation. Of the two methods, restoration of the inferomedial cortical bone support showed better dynamic and static biomechanical properties than placement of calcar screws alone.
Keywords