PLoS ONE (Jan 2024)

Reliability and minimal detectable change of the MX3 hydration testing system.

  • Ian Winter,
  • Josie Burdin,
  • Patrick B Wilson

DOI
https://doi.org/10.1371/journal.pone.0313320
Journal volume & issue
Vol. 19, no. 11
p. e0313320

Abstract

Read online

Assessing hydration status outside of laboratories can be challenging given that most hydration measures are invasive, stationary, costly, or have questionable validity. This study investigated the within-day, test-retest reliability, and minimal detectable change (MDC) of the MX3 Hydration Testing System (MX3 Diagnostics), a relatively low cost, noninvasive, and portable method to measure saliva osmolality. Seventy-five adults (44 men, 31 women; 29.6±10.8 yr, 171.1±9.2 cm, 79.1±15.4 kg) presented two saliva samples approximately 3 to 5 minutes apart. Fluid intake was avoided for at least 5 minutes prior to sample collections. For each sample collection, a researcher used the MX3 to tap the tip of a test strip to saliva on a participant's tongue. Intra-class correlation coefficients (ICCs) and MDC were calculated as measures of reliability. Means for the two measurements were 64.3±19.0 mOsm and 65.5±17.9 mOsm (mean difference of -1.2±13.0 mOsm; t = -0.806, p = 0.423). Further analysis with a two-one-sided test procedure revealed equivalence between the two SOSM measurements (p 20 mOsm) in saliva osmolality outside of laboratory constraints.