Dose-Response (Oct 2018)

Image Quality and Dose Reduction by Dual Source Computed Tomography Coronary Angiography: Protocol Comparison

  • Ernesto Forte,
  • Serena Monti,
  • Chiara Anna Parente,
  • Lukas Beyer,
  • Roberto De Rosa,
  • Teresa Infante,
  • Carlo Cavaliere,
  • Filippo Cademartiri,
  • Marco Salvatore,
  • Christian Stroszczynski,
  • Carlo Tedeschi

DOI
https://doi.org/10.1177/1559325818805838
Journal volume & issue
Vol. 16

Abstract

Read online

Purpose: To compare image quality and radiation dose among different protocols in patients who underwent a 128-slice dual source computed tomography coronary angiography (DSCT-CTCA). Methods: Ninety patients were retrospectively grouped according to heart rate (HR): 26 patients (group A) with stable HR ≤60 bpm were acquired using high pitch spiral mode (FLASH); 48 patients (group B) with irregular HR ≤60 bpm or stable HR between 60 and 70 bpm using step and shoot mode; and 16 patients (group C) with irregular HR >60 bpm or stable HR ≥70 bpm by retrospective electrocardiogram pulsing acquisition. Signal to noise ratio (SNR) and contrast to noise ratio (CNR) were measured for the main vascular structures. Moreover, the dose-length product and the effective dose were assessed. Results: Both SNR and CNR were higher in group A compared to group C (18.27 ± 0.32 vs 11.22 ± 0.50 and 16.75 ± 0.32 vs 10.17 ± 0.50; P = .001). The effective dose was lower in groups A and B (2.09 ± 1.27 mSv and 4.60 ± 2.78 mSv, respectively) compared to group C (9.61 ± 5.95 mSv) P < .0001. Conclusion: The correct selection of a low-dose, HR-matched CTCA scan protocol with a DSCT scanner provides substantial reduction of radiation exposure and better SNR and CNR.